ELEMENTRAC ONH-p 2 Oxygen / Nitrogen / Hydrogen Analyzer
ELEMENTRAC ONH-p 2 Oxygen / Nitrogen / Hydrogen Analyzer
Description
The ELEMENTRAC ONH-p determines oxygen, nitrogen, and hydrogen in inorganic samplesby inert gas fusion in an impulse furnace with temperatures in excess of 3.000 °C.
The ELEMENTRAC ONH-p guarantees precise and fast sample analysis. The analyzer covers a wide range of applications such as metal, ceramics and other inorganic materials.
The ELEMENTRAC ONH-pcan be supplied with up to two infrared cells with different path lengths, accommodating both high and low level oxygen analyses. Nitrogen and hydrogen concentrations are determined in the ELEMENTRAC ONH-pby a robust and sensitive thermal conductivity cell.
Product Advantages
Features
Measured elements: Hydrogen, Nitrogen, Oxygen
Samples: inorganic
Furnace alignment: vertical
Sample carrier: graphite crucibles
Fields of application: ceramics, engineering / electronics, steel / metallurgy
Furnace: electrode impulse furnace (max. 8,5 KW*), temperatures in excess of 3,000 °C
Detection method: solid state infrared absorption for oxygen thermal conductivity for nitrogen and hydrogen
Typical analysis time: 120 - 180 s
Chemicals required: copper oxide, magnesium perchlorate, Schuetze reagents, sodium hydroxide
Gas required: compressed air, helium 99.995 % pure, nitrogen 99.995% pure, argon 99.995% pure (if required), all gases with (2 - 4 bar / 30 - 60 psi)
Power requirements: 3~ 400 V, 50/60 Hz, max. 8,500 W
Dimensions (W x H x D): 57 x 77 x 63 cm
Weight ~ 161 kg
Required equipment: balance (resolution 0.0001g), monitor, PC
Optional accessories: carrier gas purification, external chiller, gas calibration unit